Modelling and Optimization of a Vaporizer for Inhalation Drug Therapy

, , , , , ,
,

Keywords: , , ,

We present a piezo-electrically driven droplet generator with a large number of nozzles working in parallel and operating in the Drop On Demand mode. This device is build as part of a system for Inhalation Drug Therapy. Details of modelling and optimisation of this device are explained. The operation of the device is resonance dependent. To arrive at a higher throughput and a more reliable operation, modeshape analysis has been performed. Finally an improved design was proposed which uses modeshape targeting. Final results are a fast prototyping platform and a set of design rules.

PDF of paper:


Journal: TechConnect Briefs
Volume: Technical Proceedings of the 1999 International Conference on Modeling and Simulation of Microsystems
Published: April 19, 1999
Pages: 542 - 545
Industry sector: Sensors, MEMS, Electronics
Topics: Micro & Bio Fluidics, Lab-on-Chip, Modeling & Simulation of Microsystems
ISBN: 0-9666135-4-6